Connect directly to any XR440, XR450 or XR5-SE data logger; no external power required.
P350-D-0.1-inch Pressure Sensor (±0.1 inches H2O)
Applications include Lab/Clean Room Pressurization, Filter Pressure, Fan Static Pressure.
Compatible pressure media: dry air and inert gases: nitrogen, helium, neon, argon, krypton, xenon.
Total Error Band | 2% of span from 5 to 45°C (41°F to 113°F) |
Operating Temperature Range: | -20°C to 60°C (-4°F to 140°F) |
Storage Temperature Range: | -20°C to 70°C (-4°F to 158°F) |
Resolution: | See table below |
Supply Voltage | 4.75 to 5.25 Vdc |
Power requirement: | 2 ma at 5 Vdc (supplied by Pace data logger) |
Output signal | 0.5 to 4.5 Vdc at 5Vdc supply voltage. Output is ratiometric to supply voltage. |
Wiring: | White: E terminal of Pace logger (Power) Green: Analog Input of Pace Logger (Signal) Black: "C" terminal of Pace logger (Ground) |
Proof Pressure:* | 7.5 psi |
Weight: | 1.3 oz. (37 grams) |
Physical: | Thermoplastic housing with three twisted leads, 24 AWG, 18" length. |
Pressure Port: | Port accepts 3/32" I.D. tubing (or 2.4 mm I.D. tubing). Accepts 1/8" I.D. tubing (or 3.2 mm I.D. tubing) with included adapter (details below). |
Mounting: | Two mounting slots on base, spaced 3.46" (87.8 mm) center to center. Slots are 0.138" x 0.6" (3.5 x 15 mm) |
Overall Size: | 3.79" x 1.58" x 0.88" (97 x 40 x 22 mm) |
* Exceeding Proof Pressure can result in sensor failure.
Included Adapters
Allows pressure connections to commonly available 1/8" I.D. tubing. Consists of two 8" lengths of 3/23 I.D. tubing and two 3/32" I.D. to 1/8" I.D. nylon adapter fittings. Adapters included with each P350 Pressure Sensor.
Range | Resolution inch H2O | Resolution Pascals |
---|---|---|
± 0.1 inch H2O | 0.00006 | 0.015 |
± 0.5 inch H2O | 0.0003 | 0.075 |
± 1 inch H2O | 0.0006 | 0.15 |
Part No. | Range | Price |
---|---|---|
P350-D-0.1-inch | ± 0.1 inch H2O | $180. |
P350-D-0.5-inch | ± 0.5 inch H2O | $180. |
P350-D-1-inch | ± 1 inch H2O | $180. |
P350 Sensors can be read in any pressure unit (psi, kPa, bar, etc.); see XR440 / XR450 Pressure Converter or XR5 Pressure Converter.
To read Gage pressure, leave the Low port unconnected (exposed to ambient pressure).
The 18 inch leads on all P350 Pressure Sensors may be extended up to 100 FT (30 meters) by the user using three conductor 22 or 24 gage cable. In electrically noisy environments we recommend using shielded cable.
Commonly available 1/8" I.D. or 3.2 mm I.D. plastic tubing is compatible with the P350 using the included tubing adapter.
Tubing stocked at Pace
Clear 3/32" I.D. tubing connects directly to all P350, P400, P450 and P600 Sensors. Temperature range -40°C to 74°C (-40°F to 165°F). 5/32" (4 mm) O.D., minimum bend radius 3/8" (10 mm). Maximum pressure 43 PSI. Available in 25 FT (7.6 meter) rolls.
Part Nbr: T-332-25FT Price: $10.00 25 ft (7.6 meter) roll
Multi-channel Pressure Data Logging